Combined Environmental Test Chamber Series
Environmental Reliability Testing
Combines temperature, humidity, vibration, and pressure to reproduce complex operating conditions and validate overall reliability.
High & Low Temperature / Low Pressure Test Chamber Series
High & Low Temperature / Low Pressure Test Chamber
Temperature & Low Pressure Together
-70 °C to +100 °C · ambient to 0.5 kPa (low limit of the large-capacity model: -40 °C)
Fast Pressure Reduction
Ambient to 1 kPa within 30 min (no load, dry chamber)
Smooth Pressure Recovery
Recovery rate ≤10 kPa/min
Stable Temperature Field
Fluctuation ≤±0.5 °C, deviation ≤±2.0 °C (ambient pressure, no load)
Single-factor and combined temperature-altitude testing of complete units, assemblies, components and materials under high temperature, low temperature and low pressure (altitude), for both heat-dissipating and non-heat-dissipating specimens.
Core Technical Parameters
Technical Parameters
Temperature Range
-70 °C to +100 °C (large-capacity model: -40 °C to +100 °C)
Pressure Range
Ambient to 0.5 kPa
Temperature Fluctuation
≤±0.5 °C (ambient pressure, no load)
Temperature Deviation
≤±2.0 °C (ambient pressure, no load)
Temperature Change Rate
0.7 °C to 1.0 °C/min (overall average)
Pressure Reduction Rate
Ambient to 1 kPa within 30 min (no load, dry chamber)
Pressure Recovery Rate
≤10 kPa/min
Optional Configurations
Custom Port Holes
Diameter and quantity per cable requirements
Custom Sample Rack
Designed to the specimen and fixture
Additional Shelves
Configured to specimen quantity
Product Specifications
PRODUCT PARAMETER
| Model | KU-216S | KU-504S | KU-1000S | KU-4500L | |
|---|---|---|---|---|---|
| Inner chamber dimensions (mm) | Width W | 600 | 800 | 1000 | 1500 |
| Height H | 600 | 900 | 1000 | 1740 | |
| Depth D | 600 | 700 | 1000 | 1730 | |
| Volume | 216 L | 504 L | 1000 L | 4500 L | |
| Temperature range | -70 °C to +100 °C | ||||
| Temperature fluctuation | ≤±0.5 °C (ambient pressure, no load) | ||||
| Temperature deviation | ≤±2.0 °C (ambient pressure, no load) | ||||
| Temperature change rate | 0.7 °C to 1.0 °C/min (overall average) | ||||
| Pressure range | Ambient to 0.5 kPa | ||||
| Pressure reduction rate | Ambient to 1 kPa within 30 min (no load, dry chamber) | ||||
| Pressure recovery rate | ≤10 kPa/min | ||||
| Manufacturing & Inspection Standards | JJF 1101-2019, GB/T 2424, GB/T 5170.1-2016, GB/T 5170.2-2017, GB/T 10589-2008, GB/T 10592-2008, GB/T 11158-2008, GB/T 5170.10-2017, GB/T 10590-2006, GB/T 10591-2006, GB/T 5170.17-2005 | ||||
| Compliant Test Standards | GB/T 2423-2012 (Test A cold, B dry heat, Cab damp heat steady state, Db damp heat cyclic, Z/AD, Cy 85/85), GB/T 2423.21-2008 (Test M: low air pressure), GB/T 2423.25-2008 (Test Z/AM: cold/low air pressure), GB/T 2423.26-2008 (Test Z/BM: cold/low air pressure), GJB 360B Method 103 damp heat, GJB 360B Method 106 moisture resistance, GJB 150A.3 dry heat, GJB 150A.4 cold, GJB 150A.9 damp heat | ||||
Dimensions and appearance can be customized to customer requirements; the low-temperature limit of model KU-4500L is -40 °C; specifications are subject to change without notice.
Selection Support & Consulting
Selection for low-pressure testing depends on how the temperature profile combines with the target pressure points, the specimen's size and dissipated power, and the site power and cooling-water conditions, all confirmed together with the applicable standards and test profile.
Requirement Review
Confirm the high/low temperature points and cycling profile, target pressure (altitude), specimen size and dissipated power, applicable standards and test items.
Solution Selection
Match the model by 216 L to 4500 L inner volume, temperature and pressure range, and check the cooling-capacity margin for heat-dissipating specimens.
Custom Design
Cable ports, viewing windows, sample racks, vacuum pump and pressure-control loop are configured to the specimen and test profile.
Delivery & Service
Installation and commissioning, verification of pressure reduction and recovery rates, operator training and long-term technical support.